NCCLS document EP5-A2 (ISBN ). NCCLS, West Valley Road, Suite , Wayne, Pennsylvania USA, THE NCCLS . Evaluation of Precision Performance (EP5-A2). (This feature is only available in GenEx Enterprise). Introduction. The EP5 module in Genex implements the. Buy CLSI EP5 A2 Ed. 2 () Evaluation Of Precision Performance Of Quantitative Measurement Methods from SAI Global.

Author: Samujar Grot
Country: Algeria
Language: English (Spanish)
Genre: Science
Published (Last): 4 December 2010
Pages: 467
PDF File Size: 18.79 Mb
ePub File Size: 11.66 Mb
ISBN: 174-9-79717-647-8
Downloads: 15168
Price: Free* [*Free Regsitration Required]
Uploader: Vudogor

Genex will estimate the precision in the data by the statistical methods required by the EP5-A2 document. The model assumes that all the U d are identically distributed, and similarly that the V dr are identically distributed and that the W drs are identically distributed.

The between-runs and between-days precision estimates are also given on lines 10 and We refer to the EP5-A2 document for details about the experiment protocol. Furthermore, the protocol defines a procedure for dealing with outliers. The fields under manufacturers claimed precisions allow testing against a claimed precision. This period lasts until data are obtained without operational difficulty for five days. This feature is only available in GenEx Enterprise. This period is intended for the experimenter to learn the experimental protocol, and also to put quality controls into place, and perform an initial evaluation of precision and other acceptance tests, e.

Each sample is displayed as a blue marker.

Evaluation of Precision Performance (EP5-A2)

The el5 should be separated by at least two hours, or if they take very long time to perform, one run per day is acceptable. Use the menus to the right under Columns to tell Genex which classification column that corresponds to days, and which column corresponds to runs.

Xlsi scatter plot displays the input data with the runs on the x-axis. If two or more concentrations are used, they can be handled as separate data sets, and be analyzed independently of each other in Genex. If there is only one run per day, the Runs: The random error is modeled according to the hierarchical model.

The control panel for EP5.

Page not available

Note that it is the absolute value of the variance that should be entered. Quality controls should be made regularly, based on quality control samples that are measured along with the rest of the experiment. Line 11 lcsi the same quantities as line 9, but for the total, within-laboratory, precision.

The experiment produces a data set with one or two runs per day and two replicas per run. The average of the two samples that form the run is marked with a red bar. After all fields have been set to appropriate values, press the run button to perform the analysis.

The analysis produces two windows, one scatter plot of the data and one data window with the computed precisions. The average of all measurements is shown on line 6. Furthermore, EP5 recommends that the clei is performed at two different concentrations. The measurements continue for another 15 days. If the claim is given as a percentage, it has to be translated to a variance by multiplying the average of the measured values. In total, data should be generated for 20 days, the five last days of the Protocol Familiarization plus the 15 days of the Precision Evaluation Experiment.

The EP5 module is found by selecting the Quality Ctrl tab, and clicking on the button labeled EP5, shown below as the second button from the left. The EP5 guidelines are intended both for developers of such methods and for users of methods who want eep5 verify the validity of given precision claims, or just measure their own precision.

Page not available

The SD min and SD max values form a confidence interval q2 the precision estimate, hence the confidence interval is [2. The purpose of EP5 is to evaluate cclsi precision of clinical laboratory measurement methods.

The alternative measurement protocol with one run per day, only requires the day classification column, and would have had 40 rows. The estimated between-samples variance the repeatabilityshown on line 9 is 2. EP5 also estimates the variance of the sum of the three errors, the so called within-laboratory precision.

Pe5 precision evaluation experiment is divided into cllsi different periods. EP5 estimates the repeatability, defined as the between-sample precision, i. Genex assumes that the experiment has been conducted according to the EP5-A2 instructions, producing data during 20 days with one or two runs per day.

Furthermore, EP5 allows testing of the estimated precisions against given precisions, e. Using the EP5 module in Genex. Finally, the last two lines indicate whether the claimed precisions were consistent clai the precision estimated from the measurements.

Device Familiarization Period is a period to learn the operation of equipment and preparation of samples required. The days and runs should be given as classification columns in the data input to Genex, as in the example.

Where X drs is the measurement on day d in run r on sample s. The picture only shows the first part of the data set, in total there are 80 sample rows, since the measurements are carried out for 20 days.

In A there should be two runs on each day, where each run consists of two test samples. We give below some of the main features of the EP5 guidelines, for a detailed description see the EP2-A2 document. If testing against a claim is not desired, these fields can be left blank.